Skip to main content

Specs of COV

The current specifications and features of the COV are listed below.

  • Accommodate standard hoses and fittings
  • Respiratory rate of 5 to 50 bpm
  • Tidal volume of 50 to 1000 mL
  • Ability to measure and control peak inspiratory pressure
  • Adjustable inspiratory to expiratory time ratio from 1:1 to 1:4
  • Ability to support triggered breaths
  • Provide Positive End Expiratory Pressure (PEEP)
  • Able to use ambient air or low pressure oxygen
  • Able to operate off generator power sources
  • Alarms for pressure thresholds
  • Alarms for motor timeouts and motor faults
  • Indirect measurement of disconnect from patient through under-pressure alarm
  • Indirect way of measuring apnea during breath assist (i.e. if it’s been too long since a breath has been requested, alarm sounds)